Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Comment: Migrated to Confluence 5.3
Composition Setup

Excerpt


ScanImage 3.7.1 is a service pack release, with several bugfixes and small enhancements relative to r3.7.0

Tip
titleEasy Upgrade

To upgrade from ScanImage 3.7.0, you can use use your existing INI/CFG/USR files! (To upgrade from earlier versions, new INI/CFG/USR files should be created.)

Enhancements

  • Support for PXI-based DAQ boards (e.g. PXIe-6124) & USB X series boards (USB-6356)
  • Allow scan angular range values up to 40 optical degrees
  • Support input channel voltage ranges of 0.2, 0.5, and 20V
  • Smarter beam parking to avoid large step transients on scan start
  • Configurable delay for motor steps, to allow for motor 'settling' before starting next slice in a stack
  • Configurable delay for shutter opening, to allow for shutter 'settling' before starting acquisition
  • Support for slower line periods of 16-128ms
  • Support small frame sizes (e.g. < 32 lines) without generating TIF errors
  • Support for negative photodiode voltages
  • 4-digit display for countdown/timer

Bugfixes

  • Motor operation reliability issues eliminated, or greatly reduced
  • Image data collected with Snapshot would disappear
  • Scans driven by ROI/LS/CT selections were incorrectly positioned
  • Change to fill fraction during FOCUS was producing error
  • Scan delay values were corrupted switching between bidirectional and sawtooth scanning
  • Power was being set incorrectly at end of P/z stacks when Return Home After Acq/Repeat is disabled
  • Pockels calibration now works in rejected-light case
  • # Frames/File adjustment now accounts for # Z Slices and Average Frames settings correctly
  • Zero XY and Zero Z operations no longer act like Zero XYZ
  • Recent Zoom and other scan setting changes are now correctly honored at start of LOOP acquisitions
  • TIF files where linesPerFrame is not a power-of-two are now correctly read by ImageJ
  • Discontinuous slow-dimension scan command waveform fixed, for case where Flyback on final line is selected
  • Hard error no longer occurs after some bad Pockels calibrations
  • Fix support for Matlab 2009b, which was broken for case where exported clocks are used
  • In User Functions dialog, edit operations could affect the wrong event, inadvertently
  • Save selections in Channels... dialog were being ignored
  • Crosshair was appearing doubled on ScanImage startup
  • Override Lz was being applied even when start/end-points were not defined
  • Error indicating insufficient buffering could occur when using slower imaging rates and with Disable Image Striping enabled
  • scim_openTif(): error was occurring for cases when channels selected for saving are not contiguous
  • Sporadic errors occurring during PMT Offset determination fixed
  • Position was not correctly displayed at end of Zero XY/Z/XYZ operations; now motor position is read before setting zero
  • Some header values were not saved correctly when Pockels cell feature was disabled

ScanImage 3.7.0 README (12-23-2010)

Panel
Include Page
ScanImage r3.7 README
ScanImage r3.7 README