Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Scan Amplitude = 2V (or 4V if a differential input is employed). This corresponds approximately to field-of-view scanned using Scan Amplitude = 2.5V  (or 5V with differential input) under ScanImage 3.0/3.5.
  • For this particular galvo, measured Acquisition Delay = 100?s (Scan Delay = 200?s). These values were entered and held fixed -- i.e. the linear optimization regime was employed. This Scan Delay value is well above small-angle step response specification; a better tuned scanner/servo would likely perform better.

At the specified fill fractions, the temperature remained below 50°C during 1 hour of continuous scanning. The temperature had remained stable for at least 5-10 minutes at that point.

...

Bidirectional Scanning

 

1x

2x

3x

4x

5x

0.5 ms/line

 

NONE (0.640)

NONE (0.640)  

0.787706

 

 

1 ms/line

0.819

 

 

 

 

2 ms/line

0.819 

 

 

 

 

NOTE: Values above 0.819 were not tested.

...

  • Temperature measured using Fluke 179 multimeter using Fluke 80BK-A temperature probe (Type-K thermocouple)
  • Galvo was mounted on a standard aluminum mounting block (available from Cambridge Technology) which in turn was placed on a stainless steel shelf in a well-ventilated room/area
  • Probe was fixed to exposed side of galvo can, towards the mirror, using thermal compound
  • Baseline temperature for tests was ?24.5°C
  • Optimal fill fraction fulfilling 'linear' condition was determined by eye - i.e. it was judged that ripple had sufficiently settled by the desired ramp start position