ScanImage 5 will have customization layers for both CPU and FPGA-based processing (latter for resonant scanned imaging only) which can be used to implement online or real-time analysis, including closed-loop experiments with parallel imaging and (photo)stimulus.          

Release Dates

 PrereleaseReleaseComments
ScanImage 5May 1, 2014Nov 1, 2014Recapitulate ScanImage 4 features with more flexible hardware
ScanImage 5.1Feb 1, 2015Nov 1, 2015New capabilities incorporating Founding Member requests

...

(tick) Users can refer to "Vidrio ScanImage 5" to configure a standard ScanImage 5 system with National Instruments representatives.

  • National Instruments can include extended warranty support for the hardware bundles, You will be given the option when placing the order. Please discuss with your local sales rep if you have questions.
  • ScanImage 5 will include pre-complied FPGA bitfiles - one for each of the four FPGA/Digitizer combinations above. Theis whitepaper shows how to download bitfiles to FPGA hardware.
  • The PXIe-7962R FPGA module will likely also be supported (bitfiles provided), but this will not be as extensively tested.
  • The 7961R is suited for core imaging capabilities and limited future extensibility. The 7966 supports greater future extensibility.
  • The standard PXIe-1073 chassis is a 5-slot 'remote controller' chassis connecting to the computer via a PCIe (PCI Express) card. Both PCIe card and cable are included.

 

Galvo Scanning

ScanImage 5 Galvo will support the following input digitizer options:

...

Scanner Hardware

Resonant Scanning

Imaging requires

 

...

For resonant scanned raster imaging, users require 1x each of:

Resonant Scanner: Cambridge Technologies CRS-8 or CRS-12 scanner. Includes control board.

Galvo Scanner: Cambridge Technologies 6210 or 6215 galvo system (typ. with 4-6mm mirror). Other analog-control galvos also supported.

Notes

  • Sutter can provide a scanner assembly with sound-proof housing, with accompanying electronics housing. If interested, please contact Rick Ayer at Sutter.
  • Galvo system's max slew rate can impact the frame rate achievable at larger fields-of-view. Slower slew rates require a larger number of resonant scanner line periods for galvo flyback.

Galvo Scanning

For galvo scanned raster imaging and future scan modes, users require XY galvo scanner pair. Cambridge Technologies 6210 or 6215 galvo system (typ. with 3-6mm mirrors) and other analog-control galvo systems are supported. Galvo system characteristics impact attainable line speed and performance of future scan modes.