Wiki Markup |
---|
{composition-setup |
...
} {excerpt} ScanImage 3.7.1 is a service pack release, with several bugfixes and small enhancements relative to r3.7.0 |
...
{tip | ||
:title | =Easy Upgrade | }To upgrade from ScanImage 3.7.0, you can use use your existing INI/CFG/USR files! (To upgrade from earlier versions, new INI/CFG/USR files should be created.) |
Enhancements
...
{tip} h4. Enhancements * Support for PXI-based DAQ boards (e.g. PXIe-6124) & USB X series boards (USB-6356) |
...
* Allow scan angular range values up to 40 optical degrees |
...
* Support input channel voltage ranges of 0.2, 0.5, and 20V |
...
* Smarter beam parking to avoid large step transients on scan start |
...
* Configurable delay for motor steps, to allow for motor 'settling' before starting next slice in a stack |
...
* Configurable delay for shutter opening, to allow for shutter 'settling' before starting acquisition |
...
* Support for slower line periods of 16-128ms |
...
* Support small frame sizes (e.g. < 32 lines) without generating TIF errors |
...
* Support for negative photodiode voltages |
...
* 4-digit display for countdown/timer |
...
Bugfixes
...
h4. Bugfixes * Motor operation reliability issues eliminated, or greatly reduced * Image data collected with Snapshot would disappear * Scans driven by ROI/LS/CT selections were incorrectly positioned |
...
* Change to fill fraction during FOCUS was producing error |
...
* Scan delay values were corrupted switching between bidirectional and sawtooth scanning |
...
* Power was being set incorrectly at end of P/z stacks when _Return Home After Acq/Repeat_ is disabled |
...
* Pockels calibration now works in rejected-light case |
...
* _# Frames/File_ adjustment now accounts for _# Z Slices_ and _Average Frames_ settings correctly |
...
* _Zero XY_ and _Zero Z_ operations no longer act like _Zero XYZ |
...
_ * Recent _Zoom_ and other scan setting changes are now correctly honored at start of LOOP acquisitions * TIF files where linesPerFrame is not a power-of-two are now correctly read by ImageJ |
...
* Discontinuous slow-dimension scan command waveform fixed, for case where _Flyback on final line_ is selected |
...
* Hard error no longer occurs after some bad Pockels calibrations |
...
* Fix support for Matlab 2009b, which was broken for case where exported clocks are used |
...
* In User Functions dialog, edit operations could affect the wrong event, inadvertently |
...
* Save selections in *Channels...* dialog were being ignored |
...
* Crosshair was appearing doubled on ScanImage startup |
...
* _Override Lz_ was being applied even when start/end-points were not defined |
...
* Error indicating insufficient buffering could occur when using slower imaging rates and with _Disable Image Striping |
...
_ enabled * _scim\_openTif()_: error was occurring for cases when channels selected for saving are not contiguous |
...
* Sporadic errors occurring during PMT Offset determination fixed |
...
* Position was not correctly displayed at end of Zero XY/Z/XYZ operations; now motor position is read before setting zero |
...
* Some header values were not saved correctly when Pockels cell feature was |
...
disabled h3. ScanImage 3.7.0 README (12-23-2010) |
...
Panel | |
---|---|
Include Page | {panel} {include:ScanImage r3.7READMEephus:ScanImage r3.7 README README}
{panel}
|