ScanImage 3.7.1 is a service pack release, with several bugfixes and small enhancements relative to r3.7.0
Enhancements
- Support for PXI-based DAQ boards (e.g. PXIe-6124) & USB X series boards (USB-6356)
- Allow scan angular range values up to 40 optical degrees
- Support input channel voltage ranges of 0.2, 0.5, and 20V
- Smarter beam parking to avoid large step transients on scan start
- Configurable delay for motor steps, to allow for motor 'settling' before starting next slice in a stack
- Configurable delay for shutter opening, to allow for shutter 'settling' before starting acquisition
- Support for slower line periods of 16-128ms
- Support for negative photodiode voltages
- 4-digit display for countdown/timer
- Support small frame sizes (e.g. < 32 lines) without generating TIF errors
Bugfixes
- Motor operation reliability issues eliminated, or greatly reduced
- Image data collected with Snapshot would disappear
- Scans driven by ROI/LS/CT selections were incorrectly positioned
- Change to fill fraction during FOCUS was producing error
- Scan delay values were corrupted switching between bidirectional and sawtooth scanning
- Power was being set incorrectly at end of P/z stacks when 'Return Home After Acq/Repeat' is disabled
- Pockels calibration now works in rejected-light case
- # Frames/File adjustment now accounts for # Z Slices and Average Frames settings correctly
- Zero XY and Zero Z operations no longer act like Zero XYZ
- Recent Zoom and other scan setting changes are now correctly honored at start of LOOP acquisitions
- TIF files where linesPerFrame is not a power-of-two are now correctly read by ImageJ
- Discontinuous slow-dimension scan command waveform fixed, for case where Flyback on final line is selected
- Hard error no longer occurs after some bad Pockels calibrations
- Fix support for Matlab 2009b, which was broken for case where exported clocks are used
- In User Functions dialog, edit operations could affect the wrong event, inadvertently
- Save selections in Channels... dialog were being ignored
- Crosshair was appearing doubled on ScanImage startup
- Override Lz was being applied even when start/end-points were not defined
- Error indicating insufficient buffering could occur when using slower imaging rates and with Disable Image Striping enabled
*scim_openTif(): error was occurring for cases when channels selected for saving are not contiguous
- All graphical interaction tools (line scan/ROI/power box selection) now use new functions (independent of Image Processing toolbox) with improved stability – and ability to cancel, using escape/delete.
Bugfixes
- Power Box 'OUT' value now works
- Error when parking laser for case of 2 or more beams is fixed
- 'Discard Flyback Line' option could cause errors in makeFrameByStripes() for certain scan parameters
- The 'z' hotkey (which goes to zero motor position) no longer has effect if a relative origin has not been set
- Histogram/Statistics features are working again
ScanImage 3.7.0 README (12-23-2010)
...