Model label: GalvoScanAssby

Check out the documentation for the Galvo driver box first.

Assembly instructions

 

Overview:

Y1 - Mirror with 30 degree stroke. This is the first mirror in the path of the laser. This moves in the same direction as the resonant scanner.

Y2 - Mirror with 40 degree stroke. This is the second mirror in the path of the laser.

X - This is the third, smaller galvo mirror. 

 

From left to right: Galvo driver box; in the back driver boards for X, Y1, Y2. Galvo holder with galvos and alignment jig (note pentaprism). Galvo holder with galvos and alignment jig without Y2 mirror (note nacked shaft).  

 

General:

     Set up galvo driver box.  (Note: be sure to set the input configuration for each galvo driver to full differential.  A jumper will have to be removed, as indicated in the Cambridge 671 manual.  The galvo drivers are being shipped in single-ended mode.)

     Build alignment jig. Glue pentaprism on alignment rig  

...