Emission splitting dichroic assembly: EmissSplitDichroAssby


Assembling the breadboard periscope

Creating the remote focus objective assembly

     Gluing the polarizing beam splitter to the remote focus objective assembly

     

...

     Mounting the remote focus objective     

Resonant mirror chamber

    Put cable in with the epoxy
     Assemble resonant mirror stage and actuator
     Assemble alignment jig
     Mount resonant mirror in mount using alignment jig
     Place resonant mirror into chamber
     Assembling the lid with the o-ring and tightening
     
Virtually conjugate galvo-galvo pair
     Build alignment jig
     
     Mount galvo mirrors on shafts
     
     Setup alignment laser
     Wire up and turn on galvos
     Do alignment
Assemble the V-blocks
     For pupil relay 1,2 and 3

Imaging objective mounting

     Assemble imaging objective mount
     Gluing kinematic mount
Assembling primary dichroic cap
     Putting in spring plungers and set screws
Emission splitting dichroic assembly
PMT assembly
Assemble alignment jigs
     Alignment jig cage assembly     
     Imaging objective alignment jig mount
Assembling the covers
Assembling objetive mounted optic holder
Accessory optics
1) Remove galvo mirrors from mount

EntranceQtrWavePlateAssby


1) Position the post 4" post such that its top surface is 22 mm from the top surface of the RA90, and position the 1.5" post such that it is back surface 15 mm from the back surface of the RA90.
2) Attach the the rotation slider and the quarter waveplate.Mounting galvo driver and resonant driver boards (Spencer / Steve documentation)

Assembling compressor

Assemble the retroreflector carriage
   1) Mount the AP90RL large angle bracket on the XT95P11 carriage.
   2) Attach the TSX-1D stage to the AP90RL large angle bracket.
   3) Use two M6 12mm screws to attach Thor_BA2F to retroreflector.
   4) Attach the RS1:4 to the TSX-1D stage with a set screw.
   5) Slide the BA2F onto the RS1:4. Adjust the orientation such that when the beam enters the retroreflector it will be reflected up without hitting an edge.
   

Assemble the 2x beam expander
 
 1
  1) Attach the SM1D12D ring iris to the GBE02-B 2x beam expander, making sure it is on the correct end.
   2) Attach the SM1RC lens tube rings to the TR-05:4 posts and put them in the UPH3 universal post holders.
   3) Slide the beam expander into the lens tube rings and tighten them

Assemble the prism pair compressor
Use norland 81 UV cure glue.
1) Glue prisms to TRT3 post with cutout. First glue the bottom prism, flush to the top surface of the post cutout, and positioned at a right angle. Put a few drops of UV cure glue on post and place prism on top. Do this inside a V-block to ensure the angles are correct. Once the bottom prism has been glued down, glue to top prism at a distance of 39.3mm (between outside faces). Then glue mirrors to prisms. Hold post in vice at correct orientation, place UV cure glue on prism and glue down mirror. Rotate post to glue second mirror.

2) Glue mirror and prism to polaris mirror mount, making sure orientation is correct. Attach this mirror mount to the top of the KB1X1 kinematic mount using a 8-32 screw and a nut.
3) Glue two mirrors and prisms to polaris mount and centering plate. First attach KCP2 centering plate to polaris mount. Glue one prism directly to the mount, then glue one prism to the the side of the centering plate. Leave a 1 mm space between the two prisms, and a 1 mm gap between the back face of the prism and the mount. Make sure the side of the prisms are both aligned to each other and the side of the polaris mount.
4) Attach the TRT3 post of the top plate of the KB1X1 kinematic mount using an AE8E25E internal to external thread adapter. The length of the external thread adapter needs to be shortened by about two threads to let it fit inside the TRT3 post.
5) Attach one polaris mount to the RM1F cube. Attach this cube with the RS2M spacer to the BA2 baseplate, making sure the cube is orientated such that the corner of the polaris mount is over the center of the BA2. Attach the BA2 to the Optosigma rotation stage, sliding the BA2 so the end of the slots is flush against the screws.
6) Secure the optosigma to the breadboard such that it is 25.3 mm from right side and 2.5 mm from top side.
7) Secure posts with polaris mounts to breadboard, matching the location of the items with their approximate locations in the breadboard. Ensure that the mirror, prisms all lie on the same axis.
8) Set the height of the KMCP to be 46.0 mm above the breadboard.
9) Glue custom prism onto polaris mount. Use norland 81. There are two polished and one un-polished side. Note prism is not an equilateral triangle. Make one side flush with polaris mount, and tip flush with other edge of mount.
10) Set the height of top of the iris holding posts to 87.0 mm

BeamPeriscopeAssby_2AxisStages_Solid

1) Glue a right angle mirror MRA20-P01 to polaris mount using norland 81, making side flush to bottom of mount and front flush to top of mount.
2) Glue a right angle mirror MRA20-P01 5.1 mm from edge with dial, and flush with the back of the stage.
3) Attach stages to the SolidPeriscopeBase, and attach the polaris mount to one of the stages in the correct orientation.
4) Remove redundant knob on polaris mount, as it should not be adjusted

Resonant mirror assembly: ResMirrChamber

1) Screw in schneeberger slider from underneath, make sure to use the correct 10 mm long screws so as not to damage stage
2) Add the stop screw, spring plunger, and threaded bushing (remove the lock screw) and lock nut, knob, and screw for positioning the stage.
3) Glue in the CRS resonant mirror cable using a black RTV sealant (Dow-Corning 732 multipurpose sealant), leaving 7" outside of the chamber. Let the glue cure overnight.
4) Align the resonant scanner inside its mount. UV cure a mirror to the resonant mirror alignment jig. Slide the body of the resonant scanner inside its mount and then attach the mount to the alignment jig. Adjust the height of the resonant scanner such that the center of the mirror is aligned with the top surface of the alignment jig. Rotate the resonant scanner such that its mirror is at the same angle as the alignment mirror. This angle can be determined by shining a laser beam at both mirrors simultaneously and rotating the resonant scanner inside its mount until the reflect beams are parallel.
EntranceQtrWavePlateAssby
1) Position the post 4" post such that its top surface is 22 mm from the top surface of the RA90, and position the 1.5" post such that it is back surface 15 mm from the back surface of the RA90.



Virtually conjugate galvo-galvo pair
     Build alignment jig
     
     Mount galvo mirrors on shafts
     
     Setup alignment laser

     Wire up and turn on galvos

     Do alignment



XXXXX Prepare V-blocks — insert photo of parts with screws


Imaging objective mounting

     Assemble imaging objective mount

     Gluing kinematic mount

Assembling primary dichroic cap
     Putting in spring plungers and set screws
Emission splitting dichroic assembly

PMT assembly

Assemble alignment jigs
     Alignment jig cage assembly     

     Imaging objective alignment jig mount



QQQQQ Creating the remote focus objective assembly
     Gluing the polarizing beam splitter to the remote focus objective assembly
     Mounting the quarter waveplate 

     Mounting the remote focus objective     





Assembling the covers

Assembling objetive mounted optic holder

Accessory optics
1) Remove galvo mirrors from mount


Mounting galvo driver and resonant driver boards (Spencer / Steve documentation)
2) Attach the the rotation slider and the quarter waveplate.


Making samples
     Beads
     Fluorescein 

...