Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Image Modified

Model label: ResMirrChamberAssby

Assembly instructions

1) Screw in schneeberger slider from underneath, make sure to use the correct 10 mm long screws so as not to damage stage

2) Add the stop screw, spring plunger, and threaded bushing (remove the lock screw) and lock nut, knob, and screw for positioning the stage.
3) Glue in the CRS resonant mirror cable using a black RTV sealant (Dow-Corning 732 multipurpose sealant), leaving 7" outside of the chamber. Let the glue cure overnight. Alternatively the cable can be glued after the resonant mirror is mounted since the short cable is hard to manage inside the chamber.

...

8) After mounting the mirror make sure to get all of the dust out of the chamber and seal it up with an O-ring and lid. There is no need to screw the lid down now, as it will need to be removed when the chamber is mounted to the vertical breadboard.

Alternate resonant mirror alignment instructions

Following the instructions above may not result in the resonant mirror being aligned well enough.  This alternate alignment procedure should probably be followed instead.  This procedure contains a more accurate method to check the alignment of the resonant mirror, which can be used in case the procedure above has already been followed, to check if the alternate procedure is necessary.

...