Values are color-coded as follows:

Red: Maximum safe value found (if any) is below the optimal fill fraction value determined for the linear regime (given in parentheses). In other words, thermal considerations are the limiting factor at this speed/amplitude.

Orange: Value represents the optimal fill fraction value determined for the linear regime. At that value, stable temperature was reached in the range of 40-50°C at 1 hr. Higher fill fractions, which would be in the nonlinear regime, were not tested and appear to be at risk of overheating.

Green:  Value represents the optimal fill fraction value determined for the linear regime (or 0.819, whichever is lower). At that value, stable temperature was reached below 40°C at 1 hr. Higher fill fractions were not tested, but appear to be possible.

(warning)   These results, of course, are provided as a reference only. They are in no way a guarantee.

Sawtooth Scanning


16° 1x

2x 3x

5.33°

4x 5x

3.2°

1 ms/line

NONE (0.554)

NONE (.621) 

0.585  (.640)

0.621 (.640)

0.640

2 ms/line

0.706

0.759

0.787

0.819

0.819

Bidirectional Scanning

 

1x 16°

2x 3x

5.33°

4x 5x

3.2°

0.5 ms/line

NONE (0.640)

NONE (0.640)

0.706

0.706

0.706

1 ms/line

0.819

0.819

0.819  

0.819 

0.819 

2 ms/line

0.819 

0.819

0.819

0.819

0.819