Table of Contents |
---|
...
|
Release Schedule
ScanImage 5 will be released in two parts, SI5-Resonant and SI5-Flex2014. The next version, ScanImage 2015, will add new features to ScanImage 5.
Early-Access Release(s) | Public Release | Comments | |||
---|---|---|---|---|---|
SI 5 (Resonant) | May 22 2014 (beta) | SFN 2014 | Recapitulate ScanImage 4 (resonant scanning) features with | ||
SI 2015 | SFN 2014 (beta) | SI5-Flex | SFN 2014 & | SFN 2015 | random-access imaging, and photoablation applicationsNew features beyond ScanImage 5, including galvo arb-scanning, multi-scanner, and other features specified by ScanImage 5 members. |
Early-access releases will be available exclusively to ScanImage 5 founding members.
ScanImage 5 will have customization layers for both CPU and FPGA-based processing (latter for resonant scanned imaging only) which can be used to implement online or real-time analysis, including closed-loop experiments with parallel imaging and (photo)stimulus.
SI5-Resonant Scanning Hardware
For resonant scanned raster imaging, both digitizer hardware and scanner hardware are required.
Digitizer Hardware
ScanImage 5 Resonant will use National Instruments PXI FlexRIO platform.
The system consists of three components: PXI chassis, FPGA Module, and digitizer module.
The NI system bundles with SI5 support include:
...
- The PXIe-1073 is a 5-slot 'remote controller' chassis connecting to the computer via a PCIe (PCI Express) card. Both PCIe card and cable are included.
- More advanced chassis options will work as well.
...
...
- PXIe-7975R (and other new devices) will also be supported, after August 2014. Offers greater processing power than 7961R.
In August 2014, NI will add support for NI 5732/4 digitizer modules with the PXIe-7975. - PXIe-7962R will also be supported (bitfiles provided), but not as extensively tested
- PXIe-7965/6R is NOT recommended at this time, as the PXIe-7975R will offer greater performance per unit cost.
...
NI 5732 (2 Chans/80MHz)
NI 5734 (4 Chans/120MHz)
...
...
ScanImage 2015 will support the same scanner/digitizer hardware as ScanImage 5
...
ScanImage 5 Scanner Hardware
ScanImage 5 requires a pair of scanners (one resonant and one galvo scanner) to achieve video-rate area scanning operation.
ScanImage 5 users can choose either:
scanner system that combines a resonant scanner and galvo scanner
scanner components - one resonant scanner and one galvo scanner
ScanImage 5 Scanner Systems
Sutter Instrument systems:
- Resonant Scan Box
- MOM-Resonant microscope systems
ScanImage 5 Resonant Scanner Components
Cambridge Technology components
- CRS 8Khz (part 6SC08KA012-02Y)
- CRS 12Khz (part 6SC12KA012-02Y)
- Classic CRS-8 scanner
ScanImage 5 Galvo Scanner Components
Any analog-driven galvo scanner component, including galvanometer-mounted mirror and servo driver circuit, can be used.
The following are typical choices from Cambridge Technology
Class 0 servo topology is recommended, but not required
...
ScanImage 5 DAQ Hardware
ScanImage 5 will use National Instruments PXI FlexRIO platform.
ScanImage 5 has required, user-selected, and user-optional DAQ components
Tip |
---|
Stay tuned for a National Instruments ScanImage 5 configuration tool We anticipate this may include small discounts on hardware |
User-Selected DAQ Components
Digitizer Module
FPGA Module
- Basic Option: PXIe-7961R
- Performance Option: PXIe-7975R
For SI 5, there will be no difference in operation.
Features in future versions (SI 2015 & beyond) may utilize the far
greater (~10x) image processing power of the PXIe-7975R, e.g. closed-loop experiments.
Tip | ||
---|---|---|
| ||
The PXIe-7975R option will be supported by ScanImage 5 by August 2014. Vidrio tentatively has access to a small pool of PXIe-7961R loaners Loaners are intended for users who purchase a PXIe-7975R |
User-Optional DAQ Components
PCIe-6321 option is cheaper, but requires additional wiring and uses a computer PCIe slot
HTML Comment | ||
---|---|---|
| ||
HTML Comment | ||
---|---|---|
| ||
Computer Hardware
...
Scanner Hardware
Tip | ||
---|---|---|
| ||
Sutter Instrument company offers a Resonant Scan Box component combining the required components in a sound-proof housing. Both the Sutter standalone Resonant Scan Box and their full MOM-Resonant microscope system will be compatible with ScanImage 5. |
Info | ||
---|---|---|
| ||
ScanImage 5 supports the CRS Resonant Scanner system from Cambridge Technology. Recently, Cambridge is offering a new CRS Resonant Scanner Driver board (part 6SC08KA012-02Y with new 711-80159 driver board). The new systems provide improved synchronization and are optimized for use with the maximal angle versions of the scanner devices. Vidrio engineers will test this new scanner system before April 1, 2014. Prospective ScanImage 5 users may elect to wait until this assessment before purchasing a scanner system. Note that 'classic' CRS scanner systems will also be fully supported, though minor additional electronics might be needed. |
For resonant scanned raster imaging, users require 1x each of:
Resonant Scanner: Cambridge Technologies CRS-8 or CRS-12 scanner (older or newer systems. Includes control board.
Galvo Scanner: Cambridge Technologies 6210 or 6215 galvo system (typ. with 4-6mm mirror), with servo control board. Other analog-control galvos also supported.
Notes
CRS-12 support will not be ready by the SI5-Resonant prerelease date; its support is planned by SFN 2014
Galvo system's max slew rate can impact the frame rate achievable at larger fields-of-view. Slower slew rates require a larger number of resonant scanner line periods for galvo flyback.
SI5-Flex Galvo Scanning Hardware
For SI5-Flex galvo scanning modes (raster and future scan modes):
Digitizer Hardware
SI5-Flex will support the following input digitizer options:
Family | Model Number(s) | Comments |
---|---|---|
PCIe X Series | PCIe-6351, PCIe-6353 | PCIe-6351/6353 recommended for no-Pockels/Pockels cases, respectively |
PXIe X Series | PXIe-6356, PXIe-6358 | PXIe boards fit into chassis used for resonant scanning (if applicable). PXIe-6356/6358 recommended for no-Pockels/Pockels cases, respectively. |
S Series | PCI-6110 | Not recommended for new systems, but will remain supported. A PCI slot and full-size computer case is required to house this board. |
HTML Comment | ||
---|---|---|
| ||
<div class="table-wrap"><table class="confluenceTable"><tbody><tr><td colspan="1" class="confluenceTd">USB X Series</td><td colspan="1" class="confluenceTd"><a href="http://sine.ni.com/nips/cds/view/p/lang/en/nid/209071" class="external-link" rel="nofollow">USB-6351</a>, <a href="http://sine.ni.com/nips/cds/view/p/lang/en/nid/209072" class="external-link" rel="nofollow">USB-6353</a>,<br/><a href="http://sine.ni.com/nips/cds/view/p/lang/en/nid/209077" class="external-link" rel="nofollow">USB-6356</a></td><td colspan="1" class="confluenceTd">USB-6351/6353 recommended for no-Pockels/Pockels cases, respectively</td></tr></tbody></table></div> |
Scanner Hardware
An X/Y analog-controlled galvo scanner pair is required. Cambridge Technologies 6210 or 6215 galvo system (typ. with 3-6mm mirrors) is typical. Galvo system characteristics impact attainable line speed and performance of future scan modes.
HTML Comment | ||
---|---|---|
| ||
<div class="table-wrap"><table class="confluenceTable"><tbody><tr><td colspan="1" class="confluenceTd"><p>Some users elect to set up TPLSM microscopes from a microscope hardware manufacturer. The following information is provided as a benefit to users</p><p><em>Sutter</em>: Full compatibility with ScanImage 5 resonant/galvo capabilities</p><p><em>Scientifica: </em></p>Thorlabs and Prai</td></tr></tbody></table></div> |
Computer Hardware
ScanImage 5 (resonant and galvo modules) is expected to run on standard multi-core machines (e.g. iCore7) with at least 4GB RAM and at least 4 PCIe slots.
Example tested ScanImage 5 computer systems will be provided at the prerelease dateby the early-access release date.