ASSEMBLING MODULES
MOUNTING AND ALIGNING TABLE OPTICS
MOUNTING MODULES ON THE VERTICAL BREADBOARD
IMAGING
MICROSCOPE ALIGNMENT
ELECTRONICS
PREPARING SAMPLES
APPENDIX A: Accessory optical path description
APPENDIX B: Dropbox link to Nick's photographs during the workshop
https://www.dropbox.com/sh/hilocw2n2cb8afc/AAD-j6BYuydTHK7AV1_iLdPya?dl=0
SECTION 1: INITIAL MODULE ASSMBLEY
These modules should be assembled on a bench before mounting on the vertical breadboard. Assembly of the modules can be done in any order.
Entrance beam expander: EntranceBeamExpander_4x1
Image Removed1) Mount the Thor_ACN254-050-B:1 lens in its lens tubes ensuring that its curved sides will point inwards when assembled.
Image Removed2) Assemble the cage from thorlab parts including Thor_AC254-200-B-ML mounted lens. Set the lens cage plates approximately 141mm apart. The exact position of the plates will be set during microscope assembly.
Image RemovedRemote focus mirror positioner: RFAMirrorAssemb_EquipSolLFA2004
Image Removed
Image Removed1) Gather parts for mounting the mirror on the voice coil bobbin. Clean the mirror mounting screw 90666A101_TYPE 316 SS LOW PROFILE SOCKET HEAD CAP SCREW with drop of methanol. Place small drop (30 ul) of lock-tight onto threads on the voice coil bobbin. Screw in screw, and leave overnight to set.
Image Removed2) Glue remote focus mirror onto the screw using 1-day epoxy. Fill hole on top of screw with epoxy and then mount mirror centered on the screw head (slight deviations from the center of the screw head are fine).
Image Removed3) Assemble the voice coil base and mount voice coil
Image Removed
Image Removed
Image Removed***INSERT PHOTO OF FINISHED ASSEMBLY***
Fold mirror assembly: FoldMirrorAssemb
Image Removed
Image Removed
1) Attach gimbal mount to translation stage via adaptor plate.
2) Adjust thickness of gimbal mount to 29.7 mm. The knob in the bottom corner should no longer be adjusted. Mark it with a red X to ensure it doesn't accidentally get changed during the setting of the compensators.
***INSERT PHOTO OF RED X on gimbal***
3) Cement the fold mirror to the gimbal.
***Describe gluing procedure ***
***INSERT PHOTO OF FINISHED ASSEMBLY***
Imaging objective tray: IMOTrayAssby
Image Removed
Image Removed
1) Set to of plunger to be flush with the metal surface of the tray
2) Slide dowel rod in place
3) Assemble latch, making sure orientation is correct when the latch is fully closed and open. The exact height of the latch will be set during final assembly.
Image Removed
Big detection lens and accessory optics dichroic holder: BigDetLensHolderAssby
Image Removed
Image Removed
DichroMotorBaseAssby
Image Removed
Assembling base plate, screw in bearings, make sure balls do not fall out of carridge
Glue magnets into base plate - make sure poles face in same direction **** INCLUDE PHOTO ***
Glue in rack with 5 min epoxy, push down hard so flush
MovingDichroAssby
Image RemovedAssembling slider & carriage
Motor assembly
Solder motor wires to carriage connector
Mounting of accessory dichoric / mirror (gluing to spider)
BigDetLensHolderOnlyAssby
Image Removed Mounting slider in holder
Mounting lenses (big collection lens and accessory tube lenses)
MotorizedMirrorCoverAssby
Image Removed
Assemble slider cover
Emission splitting dichroic assembly: EmissSplitDichroAssby
Image Removed
Assembling the breadboard periscope
Creating the remote focus objective assembly
Gluing the polarizing beam splitter to the remote focus objective assembly
Mounting the quarter waveplate
Mounting the remote focus objective
Resonant mirror chamber
Put cable in with the epoxy
Assemble resonant mirror stage and actuator
Assemble alignment jig
Mount resonant mirror in mount using alignment jig
Place resonant mirror into chamber
Assembling the lid with the o-ring and tightening
Virtually conjugate galvo-galvo pair
Build alignment jig
Mount galvo mirrors on shafts
Setup alignment laser
Wire up and turn on galvos
Do alignment
Assemble the V-blocks
For pupil relay 1,2 and 3
Imaging objective mounting
Assemble imaging objective mount
Gluing kinematic mount
Assembling primary dichroic cap
Putting in spring plungers and set screws
Emission splitting dichroic assembly
PMT assembly
Assemble alignment jigs
Alignment jig cage assembly
Imaging objective alignment jig mount
Assembling the covers
Assembling objetive mounted optic holder
Accessory optics
1) Remove galvo mirrors from mount
Mounting galvo driver and resonant driver boards (Spencer / Steve documentation)
Assembling compressor
Assemble the retroreflector carriage
Making samples
Beads
Fluorescein
Accessory optics
SECTION 2: MOUNT ON BREADBOARD
Mounting on the breadboard
SECTION 3: WIRING
Set up scanimage
Machine data file setting
Wiring galvos / voice coil / PMTs / etc.
SECTION 4: TABLE OPTICS ALIGNMENT
Initial table optics
Pockels cell
Shutter
2x table beam expander
Assembling the compressor
Build enclosure for table optics
SECTION 5: MICROSCOPE ALIGNMENT
Multistage periscope alignment
Quarter wave plate adjustment for multistage periscope
Alignment of beam to remote focus objective
Adjustment of quarter wave plates for maximum transmission
Alignment of RF mirror assembly
Alignment of beam to imaging objective using compensators
Adjustment of galvo voltage divider gain
SECTION 6: IMAGING
Dye pool imaging to set compensators
Dye pool imaging to set retroreflector position
Image beads and measure PSF
APPENDIX A:
Accessory optical path
...