- Minimum requirement is Thorlabs Multiphoton Essentials Kit (either MPM-SCAN or MPM-2PKIT) specified as ScanImage Compatible (very important!). Such a system will include a suitable ScanImage 4.1 Workstation.
- Thorlabs system also includes 1 Alazar ATS-9440 high-speed digitizer board and 1 National Instruments (NI) PCIe-6321 board
...
- NI board used by Thorlabs should be named 'Dev1' in the Measurement and Automation Explorer (MAX).
- NI board(s) used by ScanImage can be named as desired. One suggestion
A good practice is to use name the board(s) 'si4' /and (if applicable) 'si4-2'.
Wiring Instructions
- Use this wiring diagram
- Connection of Self Trig Source to Self Trig Destination Terminal (on all SI4 NI boards) is required
- Connection of _Frame Trig Out' from Thor ECU to Frame Clock Terminal connection (on all SI4 NI boards) is required
...
On first startup of ScanImage 4.x, you will be prompted to create or select a Machine Data File (MDF). Choose to create one, selecting desired folder location, e.g. 'C:\DATA\CONFIG\SCANIMAGE'.
...
Key things to set:
primaryDeviceID: Set to Specifies NI device name for primary ScanImage NI board, e.g. 'Dev2'
extFrameClockTerminal: Typically leave as default, 0, signifying PFI0
trigSelfTrigSourceLineID: Typically leave as default, 0, signifying P0.0
trigSelfTrigDestinationTerminal: Typically leave as default, 2, signifying PFI2
trigSelfTrigSourceDeviceID: Typically leave as default, '', signifying to use primaryDeviceID board
%%Thorlabs Devices Section
si4'. All counter channels (0-3) are utilized by ScanImage 4.1 on this board.
extFrameClockTerminal: Specifies PFI terminal (e.g. 0 for PFI0) on which Frame Clock from Thor ECU is connected to all SI 4.1 NI boards.
trigSelfTrigSourceDeviceID: Specifies NI device name (e.g. 'si4') for board on which ScanImage self-trigger signal is generated.
trigSelfTrigSourceLineID: Specifies digital output line on Port 0 (e.g. 0 for P0.0) on which ScanImage self-trigger signal is generated.
trigSelfTrigDestinationTerminal: Specifies PFI terminal (e.g. 2 for PFI2) on which all SI 4. NI boards should be connected to the ScanImage self-trigger signal source line.
%%Thorlabs Devices Section
apiCurrentVersion: Set to '1.4'. This is only valid option supported as of ScanImage 4.1.Select the API version being used – right now valid options are: '1.1' (for 32-bit/2 channel systems) and '1.2' (for 64-bit/4 channel systems)
Other Configuration
- ScanImage 4 expects the Thorlabs LSM and PMT configured XML files to be located in the same folder as the Machine Data File.
...