SECTION 1: INITIAL MODULE ASSMBLEY 

These modules should be assembled on a bench before mounting on the vertical breadboard. Assembly of the modules can be done in any order.

 

Entrance beam expander: EntranceBeamExpander_4x1

1) Mount the Thor_ACN254-050-B:1 lens in its lens tubes ensuring that its curved sides will point inwards when assembled.
2) Assemble the cage from thorlab parts including Thor_AC254-200-B-ML mounted lens. Set the lens cage plates approximately 141mm apart. The exact position of the plates will be set during microscope assembly.

Remote focus mirror positioner: RFAMirrorAssemb_EquipSolLFA2004


1) Gather parts for mounting the mirror on the voice coil bobbin. Clean the mirror mounting screw 90666A101_TYPE 316 SS LOW PROFILE SOCKET HEAD CAP SCREW with drop of methanol. Place small drop (30 ul) of lock-tight onto threads on the voice coil bobbin. Screw in screw, and leave overnight to set.
2) Glue remote focus mirror onto the screw using 1-day epoxy. Fill hole on top of screw with epoxy and then mount mirror centered on the screw head (slight deviations from the center of the screw head are fine).
3) Assemble the voice coil base and mount voice coil

***INSERT PHOTO OF FINISHED ASSEMBLY***

Fold mirror assembly: FoldMirrorAssemb


  

1) Attach gimbal mount to translation stage via adaptor plate.
2) Adjust thickness of gimbal mount to 29.7 mm. The knob in the bottom corner should no longer be adjusted. Mark it with a red X to ensure it doesn't accidentally get changed during the setting of the compensators. 
***INSERT PHOTO OF RED X on gimbal***
3) Cement the fold mirror to the gimbal.
***Describe gluing procedure ***
***INSERT PHOTO OF FINISHED ASSEMBLY***

Imaging objective tray: IMOTrayAssby

1) Set to of plunger to be flush with the metal surface of the tray

2) Slide dowel rod in place

3) Assemble latch, making sure orientation is correct when the latch is fully closed and open. The exact height of the latch will be set during final assembly.


Big detection lens and accessory optics dichroic holder: BigDetLensHolderAssby



DichroMotorBaseAssby


     
Assembling base plate, screw in bearings, make sure balls do not fall out of carridge 
Glue magnets into base plate - make sure poles face in same direction **** INCLUDE PHOTO ***     
Glue in rack with 5 min epoxy, push down hard so flush

MovingDichroAssby

Assembling slider & carriage 
          Motor assembly
    
          Solder motor wires to carriage connector
     Mounting of accessory dichoric / mirror (gluing to spider)

BigDetLensHolderOnlyAssby

     Mounting slider in holder
     Mounting lenses (big collection lens and accessory tube lenses)

MotorizedMirrorCoverAssby


     Assemble slider cover


Emission splitting dichroic assembly: EmissSplitDichroAssby




Assembling the breadboard periscope


Creating the remote focus objective assembly

     Gluing the polarizing beam splitter to the remote focus objective assembly
     Mounting the quarter waveplate 

     Mounting the remote focus objective     

Resonant mirror chamber


    Put cable in with the epoxy
     Assemble resonant mirror stage and actuator
     Assemble alignment jig
     Mount resonant mirror in mount using alignment jig
     Place resonant mirror into chamber

     Assembling the lid with the o-ring and tightening

     

Virtually conjugate galvo-galvo pair
     Build alignment jig
     
     Mount galvo mirrors on shafts
     
     Setup alignment laser

     Wire up and turn on galvos

     Do alignment

Assemble the V-blocks
     For pupil relay 1,2 and 3

Imaging objective mounting

     Assemble imaging objective mount

     Gluing kinematic mount

Assembling primary dichroic cap
     Putting in spring plungers and set screws
Emission splitting dichroic assembly

PMT assembly

Assemble alignment jigs
     Alignment jig cage assembly     

     Imaging objective alignment jig mount

Assembling the covers
Assembling objetive mounted optic holder

Accessory optics
1) Remove galvo mirrors from mount


Mounting galvo driver and resonant driver boards (Spencer / Steve documentation)

Assembling compressor

Assemble the retroreflector carriage
   1) Mount the AP90RL large angle bracket on the XT95P11 carriage.
   2) Attach the TSX-1D stage to the AP90RL large angle bracket.
   3) Use two M6 12mm screws to attach Thor_BA2F to retroreflector.
   4) Attach the RS1:4 to the TSX-1D stage with a set screw.
   5) Slide the BA2F onto the RS1:4. Adjust the orientation such that when the beam enters the retroreflector it will be reflected up without hitting an edge.
   
Assemble the 2x beam expander
   1) Attach the SM1D12D ring iris to the GBE02-B 2x beam expander, making sure it is on the correct end.
   2) Attach the SM1RC lens tube rings to the TR-05:4 posts and put them in the UPH3 universal post holders.
   3) Slide the beam expander into the lens tube rings and tighten them

Assemble the prism pair compressor
Use norland 81 UV cure glue.
1) Glue prisms to TRT3 post with cutout. First glue the bottom prism, flush to the top surface of the post cutout, and positioned at a right angle. Put a few drops of UV cure glue on post and place prism on top. Do this inside a V-block to ensure the angles are correct. Once the bottom prism has been glued down, glue to top prism at a distance of 39.3mm (between outside faces). Then glue mirrors to prisms. Hold post in vice at correct orientation, place UV cure glue on prism and glue down mirror. Rotate post to glue second mirror.

2) Glue mirror and prism to polaris mirror mount, making sure orientation is correct. Attach this mirror mount to the top of the KB1X1 kinematic mount using a 8-32 screw and a nut.
3) Glue two mirrors and prisms to polaris mount and centering plate. First attach KCP2 centering plate to polaris mount. Glue one prism directly to the mount, then glue one prism to the the side of the centering plate. Leave a 1 mm space between the two prisms, and a 1 mm gap between the back face of the prism and the mount. Make sure the side of the prisms are both aligned to each other and the side of the polaris mount.
4) Attach the TRT3 post of the top plate of the KB1X1 kinematic mount using an AE8E25E internal to external thread adapter. The length of the external thread adapter needs to be shortened by about two threads to let it fit inside the TRT3 post.
5) Attach one polaris mount to the RM1F cube. Attach this cube with the RS2M spacer to the BA2 baseplate, making sure the cube is orientated such that the corner of the polaris mount is over the center of the BA2. Attach the BA2 to the Optosigma rotation stage, sliding the BA2 so the end of the slots is flush against the screws.
6) Secure the optosigma to the breadboard such that it is 25.3 mm from right side and 2.5 mm from top side.
7) Secure posts with polaris mounts to breadboard, matching the location of the items with their approximate locations in the breadboard. Ensure that the mirror, prisms all lie on the same axis.
8) Set the height of the KMCP to be 46.0 mm above the breadboard.
9) Glue custom prism onto polaris mount. Use norland 81. There are two polished and one un-polished side. Note prism is not an equilateral triangle. Make one side flush with polaris mount, and tip flush with other edge of mount.

Making samples
     Beads
     Fluorescein 
Accessory optics

SECTION 2: MOUNT ON BREADBOARD 
Mounting on the breadboard


SECTION 3: WIRING 

Set up scanimage
     Machine data file setting
Wiring galvos / voice coil / PMTs / etc.

SECTION 4: TABLE OPTICS ALIGNMENT 
Initial table optics
     Pockels cell
     Shutter
     2x table beam expander

Assembling the compressor

Build enclosure for table optics


SECTION 5: MICROSCOPE ALIGNMENT 

Multistage periscope alignment
Quarter wave plate adjustment for multistage periscope 

Alignment of beam to remote focus objective

Adjustment of quarter wave plates for maximum transmission
Alignment of RF mirror assembly

Alignment of beam to imaging objective using compensators
Adjustment of galvo voltage divider gain

SECTION 6: IMAGING 
Dye pool imaging to set compensators

Dye pool imaging to set retroreflector position

Image beads and measure PSF


APPENDIX A: 

Accessory optical path
     Assembly, alignment, calibration, adjustment
  • No labels