7) At this point the resonant mirror can be mounted in the chamber. This is fairly tricky as it requires management of a fairly short cable and the holder needs to be mounted deep to the bottom of the resonant chamber.
Attach the cable to the mirror. Grab the resonant scanner mount with vice grip and insert into the chamber. The cable has to be trapped under the bottom of the mount - screw to the bottom of the chamber while pushing towards the alignment edge
From Left to right: Resonant mirror in its mount; holding the resonant mirror with vice grip; putting the resonant mirror into the chamber; trick to hold screws with hex key
8) 8) After mounting the mirror make sure to get all of the dust out of the chamber and seal it up with an O-ring and lid. There is no need to screw the lid down now, as it will need to be removed when the chamber is mounted to the vertical breadboard.
The following procedures will need to be completed after the modules are mounted on the microscope breadboard, and after the beam is aligned through the remote focus assembly
9) Remove