Model label: ResMirrChamberAssby

Assembly instructions

0.5) Modify the resonant mirror chamber to conform to RevA.  Use files J003086A.pdf and J003086A.stp.  This adds features that will be necessary for aligning the resonant mirror while it is in place on the microscope.  (The modification of the bottom holes to be used with flat head screws is less important than the addition of the two #8-32 holes.)

1) Screw in schneeberger slider from underneath, make sure to use the correct length screws so as not to damage stage.  Screws must not extend through static stage piece.

2) Add the stop screw, spring plunger, and threaded bushing (remove the lock screw) and lock nut, knob, and screw for positioning the stage.
3) Glue in the CRS resonant mirror cable using a black RTV sealant (Dow-Corning 732 multipurpose sealant), leaving 7" outside of the chamber. Let the glue cure overnight. Alternatively the cable can be glued after the resonant mirror is mounted since the short cable is hard to manage inside the chamber.

...

8) After mounting the mirror make sure to get all of the dust out of the chamber and seal it up with an O-ring and lid. There is no need to screw the lid down now, as it will need to be removed when the chamber is mounted to the vertical breadboard.

The following procedures will need to be completed after the modules are mounted on the microscope breadboard, and after the beam is aligned through the remote focus assembly

9) Remove