Model label: GalvoScanAssby

Assembly instructions

 

 

 

 

 

 

     Build alignment jig
     Mount galvo mirrors on shafts
     Setup alignment laser
     Wire up and turn on galvos
     Do alignment

 

1) Slide in 30 deg stroke (written on back) and 40 deg stroke (written on back) and 3rd galvo into mounting block. Push galvos all the way into mount.

2) Attach first mirror to galvo. Remove the shafts the galvos come on for shipping. Push mirror all the way along shaft.

3) Couple a test laser beam into the pentaprism, level and parallel to whole pattern on table horizontally. Use target card with target in position of pentaprism. Adjust position of whole apparatus such that the back reflection goes along the same vertical plane.

4) Hook up Y1 first, on power up make sure no faults - LED turns green. Note that X cable /  Y cables have different endings. Make sure to put mirrors on before turning galvos on.

5) Set Y 1 output voltage to 1.5V - which is approximately 6 V on the input.

6) Rotate angle of Y 1 mirror such that the reflected beam is coming of at an 8° angle.

7) Did fine adjustment of voltage offset to hit 8°, see voltage is 1.511 V

8) Mount the other Y 2 mirror, adjust position so beam perpendicular coming out -4.8mV

9) adjust X angle so comes out level, 0.6 mV

 

Recommended torques for galvo mirrors and galvos

14 mm x mirror .05” screw socket, 2 inch-pounds torque

20 mm x mirror 1/16” screw socket, 3.8 inch-pounds torque

y mirrors 140 inch-pounds torque

x mirror 110 inch-pounds torque