SECTION 2: MOUNT ON BREADBOARD
Mounting on the breadboard
Put dowel pins into holes
Mount brackets for PupilRealy3 first, make them pressed against the pins and their angled side facing down
Mount pupil relay 3, push so all the way to the left (edge against mounting surface). Tighten down clamps with screws with springs. Make sure nuts are in right place.
Mount IMOholder and tray, pushed it against the table and pins.
Mount primary dichroic, make sure caret faces down
Add primary dichro cap, pressing it against vertically mounted breadboard.
It secures to the primary dichroic holder with M4 screws. Once the cap is in place be *very* careful if you have to remove it. It's easy to chip the dichroic even if you're being careful.
Add bigDetLensHolderAssby - note need a 1/2 inch 10-24 screw to secure top right part
Add detLensCap
Mount the galvo block, pushing it against pins. Note there is not a lot of space between the galvo mirrors and pupil relay 3
Attach pupil relay 2 mount to breadboard
Use long handle quarter-20 hex drive to mount resonant chamber to breadboard. Use paper trick to get screw sitting on ball end of hex drive, and then screw chamber down. Take extra care to ensure that resonant chamber touches all three alignment pins. Check for clearance between chamber and pins using a thin (~0.001") piece of plastic (purpose-made shim stock is available, although candy wrappers, etc. can be used).
Attach pupil relay 1 mount to breadboard
Attach remote focus objective assembly
Attach fold mirror assembly
Blow all dust out of resonant mirror chamber before sealing it up
Attach emisson splitting assembly
Attach PMT assembly, with PMT cover on. Have to thread cables through cover. Screw on filters to PMT assembly.
Attach periscope
PupilRelay1 and PupilRelay2 will be attached after aligning beam through remote focus objective and using the beam to align the resonant mirror angle. To attach them, remove the resonant chamber, attach pupil relays 1&2, LOOSEN THE SCREWS on the pupil relay 1 and 2 v-block end mounts (the ones that attach them to the vertical breadboard), then remount the resonant mirror chamber and pupil relays together. Again, take extra care to check that the resonant mirror chamber is contacting the alignment pins, checking clearance using a thin shim. Tighten the screws on the pupil relay 1 and 2 v-block end mounts. Add caps to the pupil relay v-block mounts.