Overview

ScanImage requires an initialization (INI) file, usually named standard.ini, to be selected at startup. This file specifies & initializes all the state variables for ScanImage. Typically there is only one INI file per ScanImage installation (a 'rig').

ScanImage 3.5 introduces some changes to the way INI files are handled. Please familiarize yourself with the section about Initialization (INI) File Handling.

A subset of the INI file variables – termed the Rig Configuration variables – are demarcated at the beginning of the INI file. Rig Configuration variables are those which may require modification for a new ScanImage installation. These variables are rig-specific, and are not associated with either configuration (CFG) or user (USR) files. The values for these variables are not modified/modifiable during ScanImage program operation.

Rig Configuration Variable Reference

This section serves as a reference to the new and notable Rig Configuration variables in ScanImage 3.5.

Unknown macro: {report-link}

View Rig Configuration section of standard_model.ini file

.

Users should adjust the Rig Configuration values to match the hardware and experimental requirements for their rig, creating their own standard.ini file.

The standard_model.ini file contains comments which document the meaning of each of the variables.
The INI file parser is whitespace sensitive. The format <VARNAME>=<VARVALUE> must be strictly followed, i.e. no space to left or right of the equal sign

New/Notable Variables in standard_model.INI

triggerLineIndex: (Previously below Rig Configuration section) Allows specification of the static digital output line used as the shared trigger source for acquisition, mirror, and Pockels cell board(s). The use of Port 0 on the board specified by triggerBoardIndex is assumed.

triggerInputTerminal, externalTriggerInputTerminal, externalTriggerTimeout: See the sections on Triggering Simplification and External Triggering for more details.

outputBoardClockTerminal: See Clock Synchronization for more details.

maxBufferedGrabTime: See Unlimited and Uninterrupted Acquisition for more details.

maxFlybackRate: This variable is reserved for ScanImage r3.6.

scanOffsetX/Y: (Previously below Rig Configuration section) Specify offset voltages for mirror command signals. These variables are unique among the Rig Configuration variables in that they can be altered during ScanImage operation. See Beam and Scan Alignment Tools for more details.

parkAmplitudeX/Y: (Previously below Rig Configuration section) Specify voltages determining X/Y mirror positions when parked (i.e. when no acquisition is occurring). Beam safety should be considered in setting these values.

maxAmplitudeX/Y: Specify max voltage allowed for X/Y amplitudes. This clamps the sum of the scan amplitude and the scan shift set in ScanImage, for each dimension.

Shutter Variables

The separate values for 'open' and 'closed' have been removed. Variables open and epiShutterOpen are sufficient to specify the 'polarity' of shutter TTL control signal.
The variable shutterBoardIndex is new in r3.5. See Enhanced Shutter Configuration.

Motor Variables

The number of variables in the rig-specific motor section has been greatly reduced.
Two variables, firmware and posnResolution, are new in r3.5. See MP-285 Interface Improvements for more details.