Resonant mirror angle alignment issue
Issues with alignment were found when using the original resonant mirror angle alignment procedure. A new, in situ procedure was developed, which does require modifying the resonant mirror chamber part from the original design.
NEW- Custom condensor lens from IRD
The custom condensor lens (to be attached to the PMT) is not ideal. IRD was not able to manufacture the lens without a chamfer on the stem of the mushroom, which prevents the lens to fit entirely into the PMT opening. However, the gap between condensor lens and the PMT housing is sufficiently small so that we do not consider this a problem. You may have to press the lens into the PMT opening.
We were offered a 25% discount on this order, due to this issue. Please check that this discount is applied on your invoice.
Dowel pins update
Order better dowel pins: MSC (http://www.mscdirect.com/) part # 88231998 Dowel Pins;
(MCMASTER-CARR 90145A567 Dowel Pins are BAD)
Gary Zera from Zera Development will be reaching out to all of you regarding modifying two additional parts (three in total).
Alternatively, you can have your shop fix the pieces.
The surface that needs to be fixed is marked with the Diameter 3.055 dimension (zone C1). The center of this curve has to be located using the dimension 2.820 (it currently is located using the dimension 2.835).
J003103 - Primary dichroic cap
H10770PA-40 SEL - GaAsP without shuttering
H11706P-40 SEL - GaAsP with shuttering