Resonant mirror angle alignment issue

Issues with alignment were found when using the original resonant mirror angle alignment procedure.  A new, in situ procedure was developed, which does require modifying the resonant mirror chamber part from the original design.

 

NEW- Custom condensor lens from IRD

The custom condensor lens (to be attached to the PMT) is not ideal. IRD was not able to manufacture the lens without a chamfer on the stem of the mushroom, which prevents the lens to fit entirely into the PMT opening. However, the gap between condensor lens and the PMT housing is sufficiently small so that we do not consider this a problem. You may have to press the lens into the PMT opening. 

We were offered a 25% discount on this order, due to this issue.  Please check that this discount is applied on your invoice.

 

NEW- Collection of additional optional parts
These were the parts that we bought during the construction of the 2p-RAM for the workshop. In the previous partslist these were items under 'user supplied parts' (e.g. enclosure of the laser). Use your own discretion as to what you might need.
Update Aug 29: size of two mirrors have been increased.

NEW- Essential additional items for the Thorlabs kit
There are items that did not make it into the original Thorlabs kit order.

Dowel pins update

Order better dowel pins: MSC (http://www.mscdirect.com/) part # 88231998 Dowel Pins; 

(MCMASTER-CARR 90145A567 Dowel Pins are BAD)

 
Zera issues
Three pieces need fixing.

Gary Zera from Zera Development will be reaching out to all of you regarding modifying two additional parts (three in total). 

Alternatively, you can have your shop fix the pieces.

 

J003114 - Det tube

The surface that needs to be fixed is marked with the Diameter 3.055 dimension (zone C1).  The center of this curve has to be located using the dimension 2.820 (it currently is located using the dimension 2.835).

J003119a - Secondary dichroic holder 
The surfaces that need to be fixed are marked with the R.094 dimension.

J003119A.pdf


J003103 - Primary dichroic cap 

The surface that needs to be fixed is marked with the .434” dimension (zone D3 on the drawing).  See picture.  The proper datum surfaces A and B need to be used to measure this dimension correctly.

J003103.pdf





PMT issue
PMT — it looks like some of you ordered 

H10770PA-40 SEL - GaAsP without shuttering

instead of 

H11706P-40 SEL - GaAsP with shuttering

Left, H11706P-40 SE; right, H10770PA-40. 

Obviously the size of the PMTs is different. The PMT holder was designed for H11706P-40 SEL.
H10770PA-40 SEL has a smaller crossection and no mounting holes.

SOLUTION: It should be relatively easy to glue this PMT to the provided PMT holder. Simply replace the WPI Kwik-Cast with some 5 min epoxy. Make light tight with some electrical tape if necessary.

 
 

 